Test principle
In an acidic environment, inorganic phosphorus reacts with ammonium molybdate, generating ammonium phosphomolybdate complex which shows absorption at the wavelength of 340 nm. The absorbance is proportional to the concentration of inorganic phosphorus.
Product Features
1. Linear range of P level: [0.30, 4.00] mmol/L
Linear correlation coefficient ∣r∣≥0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 1.00
3. Sensitivity
The change in absorbance (ΔA) ≥ 0.01 while testing the specific P sample of concentration of 1.71 mmol/L.
4. Accuracy
Relative deviation is within ±10%.
5. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 5.0%.
Technical Parameters
Temperature |
37℃ |
Optical path length (OPL) |
1.0 cm |
Main wavelength |
340 nm |
Secondary wavelength |
700 nm (Optional) |
Method |
End-point method (single reagent) |
Reaction direction |
Positive |
Sample |
5 μL |
Reagent |
300 μL |
Reaction time |
300 s |
|
|
Reference range of P level: 0.30 – 4.00 mmol/L.
Packing Specifications
R1: 1×40 ml, R2: 2×20 ml
R1: 4×40 ml, R2: 8×20 ml
R1: 1×100 ml, R2: 2×50 ml
R1: 2×100 ml, R2: 4×50 ml
R1: 3×100 ml, R2: 6×50 ml
Storage and stability
Unopened: Store test kits at 10 - 30℃ protected from light. Stable to expiration date printed on individual vials (for 15 months).
Opened: Store test kits at 10 - 30℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.
Do not freeze the test kit.