Test principle
a-Hydroxybutyrate dehydrogenase (a-HBD) catalyzes the reversible reaction:
a-Ketobutyrate+NADH+ H+ α-HBD α-Hydroxybutyrate+NAD+
The reaction is monitored spectrophotometrically by measuring the decrease in NADH at 340 nm, and the decreasing rate of absorbance is proportional to a-HBD activity.
Product Features
1. Linear range of a-HBD activity: [3, 800] U/L
Linear correlation coefficient ∣r∣≥ 0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≥ 0.90
3. Changing rate of absorbance of blank ΔA/min ≤ 0.02
4. Sensitivity
Changing rate of absorbance of blank ΔmA/min ≥ 0.03 while testing the specific a-HBD sample of activity of 274 U/L.
5. Accuracy
Relative deviation is within ±10%.
6. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 5.0%.
Technical Parameters
Temperature |
37℃ |
Optical path length (OPL) |
1.0 cm |
Main wavelength |
340 nm |
Secondary wavelength |
380 nm (Optional) |
Method |
Velocity method |
Reaction direction |
Negative |
Sample |
5 μL |
R1 |
200 μL |
R2 |
50 μL |
Delay time |
100 s |
Reaction time |
180 s |
|
|
Reference range of a-HBD activity: 81 – 219 U/L.
Packing Specifications
R1: 1×40 ml, R2: 1×10 ml
R1: 2×40 ml, R2: 1×20 ml
R1: 2×60 ml, R2: 2×15 ml
R1: 2×80 ml, R2: 2×20 ml
R1: 4×60 ml, R2: 4×15 ml
R1: 3×80 ml, R2: 3×20 ml
Storage and stability
Unopened: Store test kits at 2 - 8℃ protected from light. Stable to expiration date printed on individual vials (for 15 months).
Opened: Store test kits at 2 - 8℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.
Do not freeze the test kit.