Test principle
Lactic dehydrogenase (LDH) catalyzes the reversible reaction:
The bidirectional reaction is monitored spectrophotometrically by measuring the increase in NADH at 340 nm in the forward reaction, and the increasing rate of absorbance is proportional to LDH activity.
Product Features
1. Linear range of LDH activity: [5, 800] U/L
Linear correlation coefficient ∣r∣≥ 0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.70
3. Changing rate of absorbance of blank ΔA/min ≤ 0.02
4. Sensitivity
Changing rate of absorbance of blank ΔmA/min ≥ 0.03 while testing the specific LDH sample of activity of 256 U/L.
5. Accuracy
Relative deviation is within ±10%.
6. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 5.0%.
Technical Parameters
Temperature |
37℃ |
Optical path length (OPL) |
1.0 cm |
Main wavelength |
340 nm |
Secondary wavelength |
450 nm (Optional) |
Method |
Velocity method |
Reaction direction |
Positive |
Sample |
10 μL |
R1 |
200 μL |
R2 |
50 μL |
Delay time |
120 s |
Reaction time |
180 s |
|
|
Reference range of LDH activity: 110 – 245 U/L.
Storage and stability
Unopened: Store test kits at 2 - 8℃ protected from light. Stable to expiration date printed on individual vials (for 15 months).
Opened: Store test kits at 2 - 8℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.