Test principle
CK-MB is the bound combination of two variants (isoenzymes CK-M and CK-B) of the enzyme phosphocreatine kinase. In the test, CKM is suppressed by sheep anti human CK-M antibody, while MK-B is free from the suppression. CK-MB level is twice as much as that of the measured CK-B.
	
Product Features
1. Linear range of CK-MB activity: [3, 800] U/L
Linear correlation coefficient ∣r∣≥ 0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.70
3. Changing rate of absorbance of blank ΔA/min ≤ 0.02
4. Sensitivity
Changing rate of absorbance of blank ΔmA/min ≥ 0.03 while testing the specific sample of activity of 58 U/L.
5. Accuracy
Relative deviation is within ±10%.
6. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 5.0%.
Technical Parameters
| 
				 Temperature  | 
			
				 37℃  | 
			
				 Optical path length (OPL)  | 
			
				 1.0 cm  | 
		
| 
				 Main wavelength  | 
			
				 340 nm  | 
			
				 Secondary wavelength  | 
			
				 410 nm (Optional)  | 
		
| 
				 Method  | 
			
				 Velocity method  | 
			
				 Reaction direction  | 
			
				 Positive  | 
		
| 
				 Sample  | 
			
				 10 μL  | 
			
				 R1  | 
			
				 200 μL  | 
		
| 
				 R2  | 
			
				 50 μL  | 
			
				 Delay time  | 
			
				 120 s  | 
		
| 
				 Reaction time  | 
			
				 180 s  | 
			
				 
  | 
			
				 
  | 
		
Reference range of CK-MB activity: 4 – 24 U/L.
	
Packing Specifications
R1: 1×40 ml, R2: 1×10 ml
R1: 2×40 ml, R2: 1×20 ml
R1: 2×60 ml, R2: 2×15 ml
R1: 2×80 ml, R2: 2×20 ml
R1: 4×60 ml, R2: 4×15 ml
R1: 3×80 ml, R2: 3×20 ml
	
Storage and stability
Unopened: Store test kits at 2 - 8℃ protected from light. Stable to expiration date printed on individual vials (for 15 months).
Opened: Store test kits at 2 - 8℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.