Test principle
The Creatine Kinase (CK) test kit is based on an enzymatic coupling method. In the reaction, phosphocreatine and ADP are concerted into creatine and ATP. The generated ATP then reacts with glucose catalyzed by hexokinase (HK), resulting in glucose-6-phosphate (G-6-P), which reacts with NADP+ to produce NADPH and 6-phospho-D-gluconate.
The concentration of CK can be calculated from the changing rate of absorbance detected at 340 nm.
Product Features
1. Linear range of CK activity: [5, 1000] U/L
Linear correlation coefficient ∣r∣≥ 0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.70
3. Changing rate of absorbance of blank ΔA/min ≤ 0.02
4. Sensitivity
Changing rate of absorbance of blank ΔmA/min ≥ 0.10 while testing the specific CK sample of activity of 327 U/L.
5. Accuracy
Relative deviation is within ±10%.
6. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 5.0%.
Technical Parameters
Temperature |
37℃ |
Optical path length (OPL) |
1.0 cm |
Main wavelength |
340 nm |
Secondary wavelength |
410 nm (Optional) |
Method |
Velocity method |
Reaction direction |
Positive |
Sample |
10 μL |
R1 |
200 μL |
R2 |
50 μL |
Delay time |
120 s |
Reaction time |
180 s |
|
|
Reference range of CK activity: 23 – 193 U/L.
Packing Specifications
R1: 1×40 ml, R2: 1×10 ml
R1: 2×40 ml, R2: 1×20 ml
R1: 2×60 ml, R2: 2×15 ml
R1: 2×80 ml, R2: 2×20 ml
R1: 4×60 ml, R2: 4×15 ml
R1: 3×80 ml, R2: 3×20 ml
Storage and stability
Unopened: Store test kits at 2 - 8℃ protected from light. Stable to expiration date printed on individual vials (for 15 months).
Opened: Store test kits at 2 - 8℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.