Cholesterol (CHOL) Test Kit (COD-PAP method)

The cholesterol (CHOL) test kit is intended for the accurate and direct quantitative measurement of CHOL in human plasma.
Brochure Download

Test principle

The test principle is shown as followings:

There is a strong absorption peak of red quinochrome at 520 nm, and the absorbance is proportional to the concentration of CHOL.  


Product Features

1.    Linear range of CHOL level: [0.20, 10.40] mmol/L

Linear correlation coefficient ∣r∣≥0.990

Linear deviation ≤ ± 10.00%

2.    Absorbance of blank ≤ 0.30  

3.    Sensitivity

The change in absorbance (ΔA) ≥ 0.01 while testing the specific CHOL sample of concentration of 5.20 mmol/L.

4.    Accuracy

Relative deviation is within ±10%.

5.    Precision

Intra CV% ≤ 5.0%; inter CV% ≤ 6.0%.

      

Technical Parameters

Temperature

37

Optical path length (OPL)

1.0 cm

Main wavelength

520 nm

Secondary wavelength

600 nm (Optional)

Method

End-point method

(single reagent)

Reaction direction

Positive

Sample

3 μL

Reagent

300 μL

Reaction time

500 s

Reference range of CHOL level: 3.10 – 6.25 mmol/L.


Packing Specifications

R1: 1×40 ml, R2: 2×20 ml

R1: 4×40 ml, R2: 8×20 ml

R1: 1×100 ml, R2: 2×50 ml

R1: 2×100 ml, R2: 4×50 ml

R1: 3×100 ml, R2: 6×50 ml


Storage and stability

Unopened: Store test kits at 2 - 8℃ protected from light. Stable to expiration date printed on individual vials (for 16 months).

Opened: Store test kits at 2 - 8℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.

Do not freeze the test kit.