Test principle
The test principle is shown as followings:
 
There is a strong absorption peak of red quinochrome at 520 nm, and the absorbance is proportional to the concentration of CHOL.
	
Product Features
1. Linear range of CHOL level: [0.20, 10.40] mmol/L
Linear correlation coefficient ∣r∣≥0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.30
3. Sensitivity
The change in absorbance (ΔA) ≥ 0.01 while testing the specific CHOL sample of concentration of 5.20 mmol/L.
4. Accuracy
Relative deviation is within ±10%.
5. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 6.0%.
Technical Parameters
| 
				 Temperature  | 
			
				 37℃  | 
			
				 Optical path length (OPL)  | 
			
				 1.0 cm  | 
		
| 
				 Main wavelength  | 
			
				 520 nm  | 
			
				 Secondary wavelength  | 
			
				 600 nm (Optional)  | 
		
| 
				 Method  | 
			
				 End-point method (single reagent)  | 
			
				 Reaction direction  | 
			
				 Positive  | 
		
| 
				 Sample  | 
			
				 3 μL  | 
			
				 Reagent  | 
			
				 300 μL  | 
		
| 
				 Reaction time  | 
			
				 500 s  | 
			
				 
  | 
			
				 
  | 
		
Reference range of CHOL level: 3.10 – 6.25 mmol/L.
	
Packing Specifications
R1: 1×40 ml, R2: 2×20 ml
R1: 4×40 ml, R2: 8×20 ml
R1: 1×100 ml, R2: 2×50 ml
R1: 2×100 ml, R2: 4×50 ml
R1: 3×100 ml, R2: 6×50 ml
	
Storage and stability
Unopened: Store test kits at 2 - 8℃ protected from light. Stable to expiration date printed on individual vials (for 16 months).
Opened: Store test kits at 2 - 8℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.
Do not freeze the test kit.