Test principle
Urea is first oxidized in the presence of uricase, generating allantoin and H2O2. H2O2 reacts further with 4-AAP and DHBS to produce a red colored product, the concentration of which is proportional to the UA level.
The test principle is shown as followings:
The increasing rate of absorbance of red quinochrome at the wavelength of 520 nm is proportional to the concentration of UA.
Performance characteristics
1. Linear range of UA level: [4, 1000] μmol/L
Linear correlation coefficient ∣r∣≥0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.40
3. Sensitivity
The change in absorbance (ΔmA) ≥ 0.10 while testing the specific UA sample of concentration of 343 μmol/L.
4. Accuracy
Relative deviation is within ±10%.
5. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 5.0%.
Technical Parameters
Temperature |
37℃ |
Optical path length (OPL) |
1.0 cm |
Main wavelength |
520 nm |
Secondary wavelength |
600 nm (Optional) |
Method |
End-point method (single reagent) |
Reaction direction |
Positive |
Sample |
4 μL |
R1 |
200 μL |
R2 |
50 μL |
Reaction time |
300 s |
Reference range of UA level: 58.7 – 415.5 μmol/L.
Packing Specifications
R1: 1×40 ml, R2: 1×10 ml
R1: 2×40 ml, R2: 1×20 ml
R1: 2×60 ml, R2: 2×15 ml
R1: 2×80 ml, R2: 2×20 ml
R1: 4×60 ml, R2: 4×15 ml
R1: 3×80 ml, R2: 3×20 ml
Storage and stability
Unopened: Store test kits at 2 - 8℃ protected from light. Stable to expiration date printed on individual vials (for 15 months).
Opened: Store test kits at 2 - 8℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.
Do not freeze the test kit.