Test principle
In specific buffer solution, positively charged ALB molecules react with negatively charged BCG molecules, forming the blue-green complex which shows absorption at the wavelength of 600 nm. The absorbance of this complex is proportional to the concentration of ALB.
ALB concentration can be calculated by comparing with the standard solution prepared in the same method mentioned above.
Product Features
1. Linear range of ALB level: [2, 60] g/L
Linear correlation coefficient ∣r∣≥0.990
Linear deviation ≤ ± 10.00%
2. Absorbance of blank ≤ 0.40
3. Sensitivity
The change in absorbance (ΔmA) ≥ 1.00 while testing the specific ALB sample of concentration of 35.3 g/L.
4. Accuracy
Relative deviation is within ±10%.
5. Precision
Intra CV% ≤ 5.0%; inter CV% ≤ 5.0%.
Technical Parameters
Temperature |
37℃ |
Optical path length (OPL) |
1.0 cm |
Main wavelength |
600 nm |
Secondary wavelength |
700 nm (Optional) |
Method |
End-point method (single reagent) |
Reaction direction |
Positive |
Sample |
5 μL |
Reagent |
300 μL |
Reaction time |
60 s |
|
|
Reference range of ALB level: 35 – 50 g/L.
Packing Specifications
R1: 1×40 ml, R2: 2×20 ml
R1: 4×40 ml, R2: 8×20 ml
R1: 1×100 ml, R2: 2×50 ml
R1: 2×100 ml, R2: 4×50 ml
R1: 3×100 ml, R2: 6×50 ml
Storage and stability
Unopened: Store test kits at 10 - 30℃ protected from light. Stable to expiration date printed on individual vials (for 15 months).
Opened: Store test kits at 10 - 30℃ protected from light. Stable for 1 month if kept capped from original container and free from contamination.
Do not freeze the test kit.